Instruction Manual CON 510 Bench Conductivity/TDS Meter 68X090820 Rev 3 Jun 07 Tech nol ogyMadeEas
Instruction Manual CON 510 Normalization Temperature: The factory default value for normalization temperature is 25 °C. If you need to normalize t
Instruction Manual CON 510 4.6 Automatic Calibration (Only for Conductivity) In the automatic calibration mode, the CON 510 bench meter is capabl
Instruction Manual CON 510 4.7 Manual Calibration (Conductivity/TDS) CALMEASoCATCREADY.uSuSCALuSoCATCMEAS.uSREADYMODEENTERRANGECALMEASCALMEASTh
Instruction Manual CON 510 4.8 Temperature Calibration oCATCCAL.oCATCCAL.uSoCATCMEASuSREADY.oCCAL...MODEENTERRANGECALMEASCALMEASMODEThe Conducti
Instruction Manual CON 510 5 MEASUREMENT The CON 510 bench meter measures up to 5 different measurement ranges with auto-ranging capability which
Instruction Manual CON 510 5.2 Manual Temperature Compensation NOTE: For manual temperature compensation, you must deactivate the ATC mode to ‘nO
Instruction Manual CON 510 5.4 Using Manual Ranging Function MEASMEASREADYATCoC.ATCoC.MEASREADYATCoCuSuSuSENTERRANGEENTERRANGEENTERRANGEAlthou
Instruction Manual CON 510 6 MEMORY AND DATA INPUT FUNCTIONS MEASoCATC.MEMuSREADY6.1 Memory Input The meter’s non-volatile memory can store and r
Instruction Manual CON 510 7 SETUP FUNCTIONS The setup mode allows you to customize the meter’s setting to your individual preferences. The CON510
Instruction Manual CON 510 SETUP . P5.0: Mode of Calibration Selection of Automatic or Manual Calibration (only in Conductivity mode) Selecti
Instruction Manual CON 510 SETUP.SETUP.SETUPK=.K=..K=.ENTERRANGECALMEAS7.3 P2.0: Viewing Electrode Diagnosis Main program 2.0 shows the effectiv
Instruction Manual CON 510 P3.2 Selecting °C or °F SETUPoC.SETUPoF. CALMEASENTERRANGEENTERRANGEYou can select between °C and °F as unit of measure
Instruction Manual CON 510 P3.4 Setting the TDS factor (In TDS Mode) The concentration of salts dissolved in solution increases the conductivit
Instruction Manual CON 510 P4.2 Adjusting the Normalization Temperature SETUPoCSETUPoC...oF...oCoFENTERRANGEENTERRANGECALMEASYour meter will
Instruction Manual CON 510 7.6 P5.0 Mode of calibration .SETUPSETUPSETUPSETUPSETUPENTERRANGEENTERRANGEENTERRANGECALMEASThe CON 510 bench m
Instruction Manual CON 510 7.7 P6.0 Selecting the cell constant SETUP.K=SETUPSETUP...SETUP..K=K=ENTERRANGEENTERRANGECALMEASThis program allows
Instruction Manual CON 510 7.8 P7.0: Resetting to factory default settings SETUPSETUPSETUP.ENTERRANGEENTERRANGECALMEASProgram 7.0 allows you to
Instruction Manual CON 510 8 PROBE CARE AND MAINTENANCE Keep the Conductivity/TDS probe clean. Rinse the probe twice, and gently swirl it while y
Instruction Manual CON 510 9 TROUBLE SHOOTING GUIDE Problem Possible Cause Solution No display A. AC outlet not switched on. A. Switch on pow
Instruction Manual CON 510 11 SPECIFICATIONS SPECIFICATIONS CON510 Conductivity Range 0 to 20.00, 200.0, 2000 µS/cm; 0 to 20.00, 200.0 mS/cm TDS
Preface This manual serves to explain the use of the CON 510 bench meter. It functions in two ways, firstly as a step by step guide to help you to op
Instruction Manual CON 510 12 ACCESSORIES Replacement Meter and Accessories Item Description Eutech Instruments Order Code No. Oakton Instrum
Instruction Manual CON 510 13 NOTE: CONDUCTIVITY AND TDS SOLUTIONS HAVE ±1% ACCURACY AT 25°C ADDENDUM 1: CALIBRATION TIPS You only need one cal
Instruction Manual CON 510 15 ADDENDUM 3: CALCULATING TEMPERATURE COEFFICIENTS To determine the temperature coefficient of your sample solution
Instruction Manual CON 510 16 ADDENDUM 4: METER FACTORY DEFAULT SETTINGS Type Parameter Default Remarks P1.1 Viewing calibration solution ––
Instruction Manual CON 510 17 WARRANTY The CON 510 bench meter is supplied with a 3-year warranty from manufacturing defects and electrodes for 6
For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distributor o
TABLE OF CONTENTS 1 INTRODUCTION 12 DISPLAY AND KEYPAD FUNCTIONS 12.1 Display 1 2.2 Keypad 2 3 PREPARATION 33.1 Conductivity Electrode I
Instruction Manual CON 510 1 INTRODUCTION Thank you for selecting the CON510 bench meter. This meter is a microprocessor-based instrument that is
Instruction Manual CON 510 2.2 Keypad The splash-proof membrane tactile keypad allows easy key entry. Each button, when pressed, has a correspondi
Instruction Manual CON 510 3 PREPARATION 3.1 Conductivity Electrode Information The CON510 bench meter is supplied with a Conductivity/TDS electr
Instruction Manual CON 510 3.4 Connecting the Electrode Holder To position the electrode arm: Use a Phillips screwdriver to remove th
Instruction Manual CON 510 4 CALIBRATION 4.1 Important Information on Meter Calibration The CON 510 bench meter allows you to perform automatic